[ Identification | Description | Examples | Input parameters | Links ]

The ILL_NeXT_Xray Instrument

NeXT (ILL) - the X-ray cone-beam tomography arm, operated simultaneously with and perpendicular to the neutron beam (companion to ILL_H521_NeXT.instr)

Identification

  • Site: ILL
  • Author: DRAFT, Peter Willendrup with Claude from public ILL / NeXT-Grenoble information
  • Origin: ILL
  • Date: September 2026

Description

DRAFT model of the X-ray imaging option of NeXT-Grenoble at the ILL.

Hardware (Tengattini et al., NIM A 968 (2020) 163939):
- source:   Hamamatsu L12161-07 sealed microfocus tube, W target, Be window,
up to 150 kV / 500 uA, minimum focal spot 5 um, 43 deg cone
- detector: Varex PaxScan 2530HE flat panel, CsI scintillator,
1792 x 2176 pixels of 139 um (portrait), up to 9 Hz (33 Hz bin 2)
- source-detector distance: >= 500 mm (set by the 43 deg cone covering the
~25 x 30 cm panel), four source positions 20 mm apart
-> SDD = 0.50, 0.52, 0.54, 0.56 m
- the sample sits on the common (neutron) rotation axis; the source-object
distance SOD sets the magnification M = SDD/SOD. The quoted 5 um
resolution needs M ~ 28, i.e. SOD ~ 18-20 mm.
- the ILL web page (2026) quotes a 20-300 kV source, i.e. the tube may have
been upgraded; kV > 150 is allowed here but flagged.

Geometry: the X-ray axis is the McXtrace z axis. At NeXT it is horizontal and
perpendicular to the neutron beam (neutrons travel along -x or +x here, depending
on the side of the source); the source/detector pivot (+-20 deg) is not modelled.

The sample defaults to the same object as in ILL_H521_NeXT: a vertical Fe rod,
radius 5 mm, for direct bimodal comparison (the neutron model gives T~0.4 at the
rod centre in white beam). An inner inclusion (e.g. an H2O-filled bore) can be added.

Detector: two ideal pixelated monitors on the flat-panel plane
Xray_counts:  photon counts (photon-counting response)
Xray_energy:  energy-weighted intensity [keV/s per pixel], a first-order model
of the energy-integrating CsI flat panel.
NOT modelled: CsI absorption efficiency vs energy, scintillator/optical blur,
focal-spot growth with tube power, heel effect, scatter from the sample
environment, and the 6 mm Pb / 10 mm + 5 mm B4C neutron shielding of the panel
(5 mm B4C over the active area transmits roughly 80-90 % at 40-150 keV).

ASSUMPTIONS (to be checked):
- Be exit window 0.2 mm (the L12161-07 datasheet value should be checked)
- focal spot = electron-beam footprint on the anode (square, focal_spot)
- anode take-off angle 12 deg (reflection-type microfocus tube, typical)
- Source_lab is a thick-target Kramers model; absolute intensity is indicative

Examples

(Test cases in bold)
  • Test: kV=150 SOD=0.1 SDD=0.5 Detector: Xray_counts_I=162607000000

Input parameters

Parameters in boldface are required; the others are optional.
NameUnitDescriptionDefault
kVkVTube acceleration voltage (<=150 for L12161-07; 20-300 per ILL web)150
tube_currentAElectron beam current (<= 500e-6)100e-6
focal_spotmFocal spot size (square electron footprint on the anode)5e-6
EminkeVLowest photon energy simulated5
anodestrAnode material file"W.txt"
take_offdegAnode take-off angle12
Be_windowmBe exit window thickness0.2e-3
filterstrPre-filter material file (e.g. "Cu.txt", "Al.txt", "Sn.txt")"Cu.txt"
filter_tmPre-filter thickness (0: no filter)0
SODmSource - object (rotation axis) distance0.1
SDDmSource - detector distance (0.50, 0.52, 0.54, 0.56)0.5
sample11: sample in beam, 0: flat field (open beam)1
sample_matstrSample (outer) material file"Fe.txt"
sample_rhog/cm3Sample density (0: take nominal density from file)0
sample_radiusmRadius of the cylindrical sample0.005
sample_heightmHeight of the cylindrical sample0.05
incl_matstrMaterial of a coaxial inner cylinder ("NULL" for none)"NULL"
incl_radiusmRadius of the inner cylinder (0 for none)0
incl_rhog/cm3Density of the inner cylinder (0: nominal)0
sample_offstrOptional OFF/PLY geometry for the outer sample, overrides the cylinder ("NULL": none)"NULL"
omegadegTomography rotation angle of the sample about the vertical axis0
det_binning1Detector binning (1: 1792x2176 native, 2, 4, ...)4

Links

  • Source code for ILL_NeXT_Xray.instr.
  • https://www.ill.eu/en/for-ill-users/instruments/instruments-list/next/
  • A. Tengattini et al., NeXT-Grenoble, the Neutron and X-ray tomograph in Grenoble,
  • Nucl. Instrum. Meth. A 968 (2020) 163939, doi:10.1016/j.nima.2020.163939
  • NECSA_MIXRAD_Nikon_Tomo.instr (McXtrace examples) - similar lab-source tomograph